Comparison of arterial pressure cardiac output monitoring with transpulmonary thermodilution in septic patients

Simon Friedrich Böttger, Dragan Pavlovic, Matthias Gründling, Michael Wendt, Orlando Hung, Dietrich Henzler, Hartmut Richard Kern, Christian Lehmann

Research output: Contribution to journalArticlepeer-review

3 Citations (Scopus)

Abstract

Background: The measurement of cardiac output in critically ill patients is complicated by rapid pathophysiological changes. The aim of this study was to compare the recently developed Arterial Pressure Cardiac Output algorithm (APCO) with transpulmonary thermodilution (TDCO). Clinical and hemodynamic parameters were tested for their impact on the measurements. Material/Methods: Twenty septic patients were examined. Cardiac output measurements were performed simultaneously on 3 consecutive days. The data were evaluated using regression analysis and the Bland Altman approach. Results: Bland Altman analysis presented a bias of 0.72 L/min and limits of agreement of -2.16 to 3.61 L/min for TDCO vs. APCO. Statistically significant covariables in the regression analysis were systemic vascular resistance (p<0.001), mean arterial pressure (p<0.001), cardiac function index (p=0.01), global end-diastolic index (p=0.02) and stroke volume index (p=0.005). Multiple linear regression analysis showed the residual percentage error decreased from 49.1% to 21.5%. Conclusions: The APCO algorithm provides a broad range of hemodynamic measurements with a minimally invasive approach and simple access to the patient's hemodynamic state. However, an underestimation at high cardiac output and an overestimation at low cardiac output relative to transpulmonary thermodilution were observed in septic patients. Therefore, the APCO algorithm in its current state cannot be substituted for transpulmonary thermodilution.

Original languageEnglish
Pages (from-to)PR1-PR7
JournalMedical Science Monitor
Volume16
Issue number3
Publication statusPublished - Mar 2010

ASJC Scopus Subject Areas

  • General Medicine

Fingerprint

Dive into the research topics of 'Comparison of arterial pressure cardiac output monitoring with transpulmonary thermodilution in septic patients'. Together they form a unique fingerprint.

Cite this