Abstract
A single-element, 40-MHz, 3-mm diameter transducer was fabricated with a geometric focus at 9 mm. The transducer was based on a piezo-composite substrate with triangular-shaped composite pillars. The 2-way bandwidth of 50% and impedance magnitude were in agreement with that predicted using finite-element modeling. A one-way radiation pattern was collected using a needle hydrophone. The one-way - 3 dB beamwidth at the geometric focus was measured to be 120 (im and the -3 dB depth of field was 2.5 mm. This is in good agreement with the theoretical predictions of 112.5 (im and 2.4 mm. The triangular-pillar composite transducer was then compared with a transducer with square composite pillars with similar volume fraction of active ceramic. A 9.5 dB reduction in the amplitude of the secondary resonance was found for the triangular-pillar composite as well as a 30% gain in the 2-way pulse bandwidth. A 256-element 30-MHz linear array was fabricated as a preliminary investigation into the use of the triangular pillar as the substrate in a high-frequency linear array transducer. In vivo images were generated with both the single-element and linear-array transducers.
Original language | English |
---|---|
Article number | 4815313 |
Pages (from-to) | 827-836 |
Number of pages | 10 |
Journal | IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control |
Volume | 56 |
Issue number | 4 |
DOIs | |
Publication status | Published - Apr 2009 |
ASJC Scopus Subject Areas
- Instrumentation
- Acoustics and Ultrasonics
- Electrical and Electronic Engineering