The effect of sterilisation on the light intensity of the reusable Trachlight

A. D. Milne, A. L. Cave, P. A. Brousseau, O. R. Hung

Research output: Contribution to journalArticlepeer-review

2 Citations (Scopus)

Abstract

Summary We examined the light intensity delivered by the Trachlight reusable lightwand and the effect of repeated sterilisation on this light intensity. Using a light meter and test chamber, the light intensity of 12 new Trachlights and 31 Trachlights available in operating theatres was measured. In addition, five new devices underwent 30 repeated sterilisations with light intensity measurement after each cleaning. There was no difference between the light intensity of new Trachlights (median (IQR [range]) 2282 (2202-2780 [1970-4400])) lux and the devices currently in service (2295 (2018-2796 [1649-3280]) lux (p = 0.27)). The median (range) number of uses for devices currently in service was only 3 (0-9). There was, however, significant light intensity loss after 10, 20 and 30 cleanings (p = 0.001). After 10 cleanings, the mean light intensity was 75% of initial output, and after 30 cycles, light intensity was 59% of initial output. Anaesthesia

Original languageEnglish
Pages (from-to)40-42
Number of pages3
JournalAnaesthesia
Volume67
Issue number1
DOIs
Publication statusPublished - Jan 2012

ASJC Scopus Subject Areas

  • Anesthesiology and Pain Medicine

PubMed: MeSH publication types

  • Journal Article

Fingerprint

Dive into the research topics of 'The effect of sterilisation on the light intensity of the reusable Trachlight '. Together they form a unique fingerprint.

Cite this