TY - JOUR
T1 - Thermostatic tissue platform for intravital microscopy
T2 - 'The hanging drop' model
AU - Pavlovic, Dragan
AU - Frieling, Helge
AU - Lauer, Kai Stephan
AU - Bac, Vo Hoai
AU - Richter, Joern
AU - Wendt, Michael
AU - Lehmann, Christian
AU - Usichenko, Taras
AU - Meissner, Konrad
AU - Gruendling, Matthias
PY - 2006/11
Y1 - 2006/11
N2 - Intravital microscopy imposes the particular problem of the combined control of the body temperature of the animal and the local temperature of the observed organ or tissues. We constructed and tested, in the rat ileum microcirculation preparation, a new organ-support platform. The platform consisted of an organ bath filled with physiological solution, and contained a suction tube, a superfusion tube, an intestine-support hand that was attached to a micromanipulator and a thermometer probe. To cover the intestine we used a cover glass plate with a plastic ring glued on its upper surface. After a routine procedure (anaesthesia, monitoring and surgery), the intestine segment (2-3 cm long) was gently exteriorized and placed on the 'hand' of the organ support. A small part of the intestine formed a small 'island' in the bath that was filled with physiological salt solution. The cover glass was secured in place. The physiological salt solution from the superfusion tube, which was pointed to the lower surface of the cover glass, formed a 'hanging drop'. The objective of the microscope was then immersed into distilled water that was formed by the cover glass plastic ring. The 'hanging drop' technique prevented any tissue quenching, ensured undisturbed microcirculation, provided for stable temperature and humidity, and permitted a clear visual field.
AB - Intravital microscopy imposes the particular problem of the combined control of the body temperature of the animal and the local temperature of the observed organ or tissues. We constructed and tested, in the rat ileum microcirculation preparation, a new organ-support platform. The platform consisted of an organ bath filled with physiological solution, and contained a suction tube, a superfusion tube, an intestine-support hand that was attached to a micromanipulator and a thermometer probe. To cover the intestine we used a cover glass plate with a plastic ring glued on its upper surface. After a routine procedure (anaesthesia, monitoring and surgery), the intestine segment (2-3 cm long) was gently exteriorized and placed on the 'hand' of the organ support. A small part of the intestine formed a small 'island' in the bath that was filled with physiological salt solution. The cover glass was secured in place. The physiological salt solution from the superfusion tube, which was pointed to the lower surface of the cover glass, formed a 'hanging drop'. The objective of the microscope was then immersed into distilled water that was formed by the cover glass plastic ring. The 'hanging drop' technique prevented any tissue quenching, ensured undisturbed microcirculation, provided for stable temperature and humidity, and permitted a clear visual field.
UR - http://www.scopus.com/inward/record.url?scp=33845769848&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=33845769848&partnerID=8YFLogxK
U2 - 10.1111/j.1365-2818.2006.01665.x
DO - 10.1111/j.1365-2818.2006.01665.x
M3 - Article
C2 - 17204068
AN - SCOPUS:33845769848
SN - 0022-2720
VL - 224
SP - 203
EP - 210
JO - Journal of Microscopy
JF - Journal of Microscopy
IS - 2
ER -