Validation study of PulseCO system for continuous cardiac output measurement

George Berberian, T. Alexander Quinn, Deon W. Vigilance, David Y. Park, Santos E. Cabreriza, Lauren J. Curtis, Henry M. Spotnitz

Research output: Contribution to journalArticlepeer-review

23 Citations (Scopus)

Abstract

Ultrasonic flow probes have been used to optimize biventricular pacing immediately after cardiopulmonary bypass, improving cardiac output (CO) by 10%; however, flow probes must be removed with chest closure. The PulseCO system (LiDCO Limited, Cambridge, UK) may extend optimization into the postoperative period, but controlled validations have not been reported. Six anesthetized pigs were instrumented for right heart bypass. Flow was varied from 3 to 1 L/min and then back to 3 in 0.5 L/min increments for 60 second intervals. CO was measured by ultrasonic flow probe on the aorta and by PulseCO using a femoral arterial line. PulseCO and flow probe accurately measured CO (PulseCO R 2: 0.79-0.95; flow probe R 2: 0.96-0.99). At flow of 2 L/min, when the heart was paced 30 bpm over the sinus rate, PulseCO falsely indicated an increase in CO (2.13 vs. 2.30 L/min, p = 0.014). When mean arterial pressure was increased by 20% using a phenylephrine infusion, PulseCO falsely indicated an increase in CO (2.13 vs. 2.47 L/min, p = 0.014). When mean arterial pressure was decreased by 20% using a nitroprusside infusion, PulseCO falsely indicated a decrease in CO (2.13 vs. 1.79 L/min, p = 0.003). PulseCO appears to be useful for assessing acute changes in CO if its limitations are recognized.

Original languageEnglish
Pages (from-to)37-40
Number of pages4
JournalASAIO Journal
Volume51
Issue number1
DOIs
Publication statusPublished - Jan 2005
Externally publishedYes

ASJC Scopus Subject Areas

  • Biophysics
  • Bioengineering
  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Validation study of PulseCO system for continuous cardiac output measurement'. Together they form a unique fingerprint.

Cite this

Berberian, G., Quinn, T. A., Vigilance, D. W., Park, D. Y., Cabreriza, S. E., Curtis, L. J., & Spotnitz, H. M. (2005). Validation study of PulseCO system for continuous cardiac output measurement. ASAIO Journal, 51(1), 37-40. https://doi.org/10.1097/01.MAT.0000150329.11072.C2