Clinical evaluation of the EMIT® salicylic acid assay

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

1 Cita (Scopus)

Resumen

Salicylic acid was measured by a homogeneous enzyme immunoassay technique (EMIT®). Replicate analyses of samples in the therapeutic range gave a within-day coefficient of variation (CV) of 5.1% and day-to-day CV of <5.0%. Quantitative results obtained by immunoassay on serum samples from patients receiving acetylsalicylic acid were compared with the results obtained by the colorimetrie procedure of Trinder. Regression analysis of 100 specimens analyzed by immunoassay (y) and the reference procedure {x) gave the equation y = 0.99x - 1.18 (r = 0.995, Syx = 1.23). Clinical evaluation of the results indicates that the enzyme immunoassay procedure is highly sensitive and specific for salicylic acid.

Idioma originalEnglish
Páginas (desde-hasta)331-334
Número de páginas4
PublicaciónTherapeutic Drug Monitoring
Volumen5
N.º3
DOI
EstadoPublished - sep. 1983

ASJC Scopus Subject Areas

  • Pharmacology
  • Pharmacology (medical)

Huella

Profundice en los temas de investigación de 'Clinical evaluation of the EMIT® salicylic acid assay'. En conjunto forman una huella única.

Citar esto