How does prosthesis head size affect vibration transmission in ossiculoplasty?

Manohar Bance, Alfredo Campos, Lillian Wong, David P. Morris, Rene van Wijhe

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

17 Citas (Scopus)

Resumen

Objectives: The transmission of vibrations from the tympanic membrane to the stapes footplate by an ossicular reconstruction prosthesis is affected by the size of the prosthesis head. We sought to determine if augmenting or reducing the head size of prosthesis had a systematic effect on transmission of vibrations to the stapes. Study Design: We conducted a fresh cadaveric temporal bone middle ear study. Methods: The incus was replaced with a prosthesis using a tympanic membrane to stapes head (TASH)-type hydroxyapatite prosthesis in nine fresh cadaveric temporal bones. Three prosthesis head sizes were created: unaltered, reduced, and augmented. Stapes vibrations were measured with a laser Doppler vibrometer in response to acoustic frequency chirps at 90 dB SPL. Results: All three head size prostheses resulted in smaller stapes vibrations than the intact ear. There was no difference in the vibration transmission between the three different head sizes. All prostheses showed a vibration loss of 10 to 15 dB compared to the intact ear. Conclusions and Significance: Within the range of sizes tested, prosthesis head size had little impact on vibration transmission to the stapes footplate.

Idioma originalEnglish
Páginas (desde-hasta)70-73
Número de páginas4
PublicaciónOtolaryngology - Head and Neck Surgery
Volumen137
N.º1
DOI
EstadoPublished - jul. 2007

Nota bibliográfica

Funding Information:
This research was partly funded by the Canadian Institutes of Health Research, Dalhousie Medical Research Foundation, and the Capital Health Research Fund.

ASJC Scopus Subject Areas

  • Surgery
  • Otorhinolaryngology