Monte Carlo Simulation of the Siemens Biograph Vision PET with Extended Axial Field of View Using Sparse Detector Module Rings Configuration

Sara A. Zein, Nicolas A. Karakatsanis, Maurizio Conti, Sadek A. Nehmeh

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

17 Citas (Scopus)

Resumen

We report on the NEMA-NU2-2012 performance of a hypothetical Monte Carlo (MC) model, Ex-PET, of the Siemens Biograph Vision positron emission tomography (PET)/CT (Bio-Vis) with sparse detector module rings and extended axial field of view (AFOV). MC simulations were performed with the detector module rings interleaved with 32-mm gaps, equivalent to the axial dimension of each detector module, yielding an AFOV of 48.0 cm (Bio-Vis has 25.6-cm AFOV). 3D-PET acquisition combined with a limited continuous-bed-motion (limited-CBM) was used to compensate for the loss in sensitivity within the gaps' regions. MC simulations of the Bio-Vis were performed for comparison purposes. All MC simulations were performed using GATE MC toolkit. Ex-PET exhibited 0.49, 0.16, and 0.16 mm deterioration in axial resolution at 1, 10, and 20 cm off-center of the transaxial field of view, respectively, compared to Bio-Vis. Only 1% reduction in system sensitivity and 6% reduction in peak NECR was observed with Ex-PET compared to Bio-Vis. 3D-OSEM image reconstruction, combined with CBM, allowed compensating for the lack of counts within the gaps' regions. NEMA Image Quality test showed < 6% reduction in contrast recovery with Ex-PET versus Bio-Vis, yet the background variability was increased by up to 8%. The feasibility of PET imaging with an easily adoptable sparse detector configuration was demonstrated. This can lay the pathway for future development of cost-effective PET systems with long and conventional AFOV's.

Idioma originalEnglish
Número de artículo9244158
Páginas (desde-hasta)331-342
Número de páginas12
PublicaciónIEEE Transactions on Radiation and Plasma Medical Sciences
Volumen5
N.º3
DOI
EstadoPublished - may. 2021
Publicado de forma externa

Nota bibliográfica

Publisher Copyright:
© 2017 IEEE.

ASJC Scopus Subject Areas

  • Atomic and Molecular Physics, and Optics
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Huella

Profundice en los temas de investigación de 'Monte Carlo Simulation of the Siemens Biograph Vision PET with Extended Axial Field of View Using Sparse Detector Module Rings Configuration'. En conjunto forman una huella única.

Citar esto