Resumen
Clinical circumstances often demand flexibility in electrodiagnostic procedures. We present a clinical case that required alteration of our routine full-field electroretinography technique. The patient presented with reports of allergy to Xylocaine® (lidocaine) and refused the use of any type of topical anesthetic drops. This conflicted with our routine ERG method that combines the use of the Henkes Lovac™ hard contact lens and the topical anesthetic Alcaine® (proparacaine hydrochloride). We acquired a successful electroretinogram without topical anesthesia by using a "piggy-back" system involving the interposition of a soft hydrophilic contact lens between the cornea and the recording hard lens. The procedure yielded excellent ERG recordings and was well tolerated. This non-routine technique was also compared to our standard ERG technique using a normal volunteer. It appears that this proposed "piggy-back" technique is a useful alternative to routine ERG procedure in cases of refusal/allergy to ophthalmic solutions and also potentially in cases where additional corneal protection is needed.
Idioma original | English |
---|---|
Páginas (desde-hasta) | 245-249 |
Número de páginas | 5 |
Publicación | Documenta Ophthalmologica |
Volumen | 116 |
N.º | 3 |
DOI | |
Estado | Published - may. 2008 |
ASJC Scopus Subject Areas
- Ophthalmology
- Sensory Systems
- Physiology (medical)