Resumen
The performance of a high frequency annular array based imaging system is described. Radiation patterns generated by the array + beamformer were evaluated by imaging a point target immersed in a water bath. The lateral image resolution at a depth of 5.5 mm was 96 microns and the secondary lobes were suppressed 60 dB below the main lobe. Preliminary images of CD-1 mice were also generated using the array system. The images were displayed with a dynamic range of up to 60 dB.
Idioma original | English |
---|---|
Título de la publicación alojada | Proceedings - 2004 IEEE Ultrasonics Symposium |
Subtítulo de la publicación alojada | A Conference of the IEEE International Ultrasonics, Ferroelectrics, and Frequency Control Society, UFFC-S |
Editores | M.P. Yuhas |
Páginas | 760-763 |
Número de páginas | 4 |
DOI | |
Estado | Published - 2004 |
Publicado de forma externa | Sí |
Evento | 2004 IEEE Ultrasonics Symposium - Montreal, Que., Canada Duración: ago. 23 2004 → ago. 27 2004 |
Serie de la publicación
Nombre | Proceedings - IEEE Ultrasonics Symposium |
---|---|
Volumen | 1 |
ISSN (versión impresa) | 1051-0117 |
Conference
Conference | 2004 IEEE Ultrasonics Symposium |
---|---|
País/Territorio | Canada |
Ciudad | Montreal, Que. |
Período | 8/23/04 → 8/27/04 |
ASJC Scopus Subject Areas
- Acoustics and Ultrasonics
Huella
Profundice en los temas de investigación de 'Performance of a 50 MHz annular array based imaging system'. En conjunto forman una huella única.Citar esto
Brown, J. A., Morton-Démoré, C., Foster, F. S., & Lockwood, G. R. (2004). Performance of a 50 MHz annular array based imaging system. En M. P. Yuhas (Ed.), Proceedings - 2004 IEEE Ultrasonics Symposium: A Conference of the IEEE International Ultrasonics, Ferroelectrics, and Frequency Control Society, UFFC-S (pp. 760-763). Artículo U2-G-4 (Proceedings - IEEE Ultrasonics Symposium; Vol. 1). https://doi.org/10.1109/ULTSYM.2004.1417833