The bovine pericardial xenograft: I. Effect of fixation in aldehydes without constraint on the tensile viscoelastic properties of bovine pericardium

J. Michael Lee, Sean A. Haberer, Derek R. Boughner

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

98 Citas (Scopus)

Resumen

Tensile testing of tissue strips has been used to examine the effect of simple fixation in glutaraldehyde and formaldehyde on the viscoelastic properties of bovine pericardium. To assess tissue anisotropy, tissue strips were cut at 0°, 30°, 60°, and 90° relative to the base‐to‐apex direction. Fresh anterior pericardium was modestly anisotropic, being least extensible in the base‐to‐apex direction; however, fixation removed this anisotropy. Fixation also produced a marked change in the response of the material to initial cyclic loading during preconditioning. Overall extensibility of the fixed material was significantly greater than that for the fresh tissue, consistent with a 10.7% shrinkage in aldehydes calculated from strain at fracture data. Reductions in stress relaxation and creep after fixation were noted as well, consistent with intrafibrillar crosslinking. Cyclic hysteresis and ultimate tensile strength were unaffected. Since the observed changes in the stress–strain response were largely attributable to shrinkage, control of shrinkage by physical means would allow for engineering modification of bovine pericardial mechanics for controlled anisotropy.

Idioma originalEnglish
Páginas (desde-hasta)457-475
Número de páginas19
PublicaciónJournal of Biomedical Materials Research
Volumen23
N.º5
DOI
EstadoPublished - may. 1989
Publicado de forma externa

ASJC Scopus Subject Areas

  • Biomaterials
  • Biomedical Engineering

Huella

Profundice en los temas de investigación de 'The bovine pericardial xenograft: I. Effect of fixation in aldehydes without constraint on the tensile viscoelastic properties of bovine pericardium'. En conjunto forman una huella única.

Citar esto