The design of ultrasonic lead magnesium niobate-lead titanate (PMN-PT) composite transducers for power and signal delivery to implanted hearing aids

Producción científica: Contribución a una revistaArtículo de la conferenciarevisión exhaustiva

10 Citas (Scopus)

Resumen

We present a system for efficiently powering implanted hearing aids by transmitting an ultrasonic signal across the skin. The use of ultrasound as method for power and signal transfer is known for embedded systems in industrial applications, and has more recently been investigated for use with other medical implants. In our application, ultrasonic transducers are investigated as they offer substantially reduced size relative to traditional magnetic induction coil power delivery. The developed transducers use PMN-PT (lead magnesium niobate-lead titanate) piezoelectric material in a 1-3 composite formulation. PMN-PT offers an electromechanical coupling factor (kt, an indicator of maximum efficiency) that is up to 60% greater than traditional piezoceramics, while the use of composite transducers removes geometric constraints that can limit the achieved efficiency. The fabrication methods for the transducers are detailed. Experimental results are presented to show the composite transducers achieve a kt of 0.77 (out of 1.00), and a power transmission efficiency of 45%.

Idioma originalEnglish
Número de artículo030029
PublicaciónProceedings of Meetings on Acoustics
Volumen19
DOI
EstadoPublished - 2013
Evento21st International Congress on Acoustics, ICA 2013 - 165th Meeting of the Acoustical Society of America - Montreal, QC, Canada
Duración: jun. 2 2013jun. 7 2013

ASJC Scopus Subject Areas

  • Acoustics and Ultrasonics

Huella

Profundice en los temas de investigación de 'The design of ultrasonic lead magnesium niobate-lead titanate (PMN-PT) composite transducers for power and signal delivery to implanted hearing aids'. En conjunto forman una huella única.

Citar esto