An Adaptive “Wiener” Filter for Estimating the Time-Derivative of the Left Ventricular Pressure Signal

S. Periyalwar, A. E. Marble, S. T. Nugent, D. N. Swingler

Résultat de recherche: Articleexamen par les pairs

3 Citations (Scopus)

Résumé

This communication presents the development of an optimal digital differentiating filter based on the Wiener theory for estimating the peak derivative of the left ventricular pressure (LVP) signal. The magnitude coherence function is used to estimate the signal and noise spectra. The peak derivative obtained by this method is found to be within 2% of the results obtained by a seven-point second-order data fit and a DFT technique.

Langue d'origineEnglish
Pages (de-à)417-420
Nombre de pages4
JournalIEEE Transactions on Biomedical Engineering
Volume37
Numéro de publication4
DOI
Statut de publicationPublished - 1990

ASJC Scopus Subject Areas

  • Biomedical Engineering

PubMed: MeSH publication types

  • Journal Article

Empreinte numérique

Plonger dans les sujets de recherche 'An Adaptive “Wiener” Filter for Estimating the Time-Derivative of the Left Ventricular Pressure Signal'. Ensemble, ils forment une empreinte numérique unique.

Citer