Changes in the functional flexion axis of the knee before and after total knee arthroplasty using a navigation system

David A.J. Wilson, Janie L. Astephen Wilson, Glen Richardson, Michael J. Dunbar

Résultat de recherche: Articleexamen par les pairs

11 Citations (Scopus)

Résumé

Long term satisfaction of patients with total knee arthroplasty (TKA) has lagged behind that of total hip arthroplasty. One possible reason is the failure of the artificial joint to recreate natural kinematics of the knee. This study evaluated the pre and post implant functional flexion axis in the knees of 285 total knee arthroplasty patients using a surgical navigation system. Results showed that post-implant there was less femoral rollback early in flexion on the lateral side of the joint than pre-implant. Designing future generations of knee implants to allow for this motion may give patients a more 'natural' feeling knee and may benefit outcomes.

Langue d'origineEnglish
Pages (de-à)1388-1393
Nombre de pages6
JournalJournal of Arthroplasty
Volume29
Numéro de publication7
DOI
Statut de publicationPublished - juill. 2014

ASJC Scopus Subject Areas

  • Orthopedics and Sports Medicine

Empreinte numérique

Plonger dans les sujets de recherche 'Changes in the functional flexion axis of the knee before and after total knee arthroplasty using a navigation system'. Ensemble, ils forment une empreinte numérique unique.

Citer