Design and fabrication of annular arrays for high-frequency ultrasound

Jeremy A. Brown, Christine E.M. Démoré, Geoffrey R. Lockwood

Résultat de recherche: Articleexamen par les pairs

84 Citations (Scopus)

Résumé

The design, fabrication, and performance of miniature high-frequency annular arrays are described. A 50-MHz, 2-mm-diameter, 7-element, equal-area annular array was fabricated and tested. The array elements were defined using photolithography and the electrical contacts were made using ultrasonic wire bonding. The resulting transducer produced pulses with a -6 dB bandwidth of 52% and an insertion loss of -16 dB. A radiation pattern was collected by scanning the transducer array above the tip of a glass fiber. A -6 dB two-way beam width of 75 microns was found at f/2. The radiation pattern decreased smoothly to less than -60 dB at a distance of 550 microns.

Langue d'origineEnglish
Pages (de-à)1010-1017
Nombre de pages8
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Volume51
Numéro de publication8
DOI
Statut de publicationPublished - août 2004
Publié à l'externeOui

ASJC Scopus Subject Areas

  • Instrumentation
  • Acoustics and Ultrasonics
  • Electrical and Electronic Engineering

Empreinte numérique

Plonger dans les sujets de recherche 'Design and fabrication of annular arrays for high-frequency ultrasound'. Ensemble, ils forment une empreinte numérique unique.

Citer

Brown, J. A., Démoré, C. E. M., & Lockwood, G. R. (2004). Design and fabrication of annular arrays for high-frequency ultrasound. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 51(8), 1010-1017. https://doi.org/10.1109/TUFFC.2004.1324405