Preclinical Evaluation of the Stealth Autoguide Robotic Guidance Device for Stereotactic Cranial Surgery: A Human Cadaveric Study

David Brandman, Murray Hong, David B. Clarke

Résultat de recherche: Articleexamen par les pairs

16 Citations (Scopus)

Résumé

Introduction: Stereotactic procedures are routinely performed for brain biopsies, deep brain stimulation, and placement of stereoelectroencephalography (SEEG) electrodes for epilepsy. The recently developed Stealth Autoguide (Medtronic, Minneapolis, MN, USA) device does not require patients to don a stereotactic frame. In this preclinical study, we sought to quantitatively compare the Stealth Autoguide robotic system to 2 devices commonly used in clinical practice: the Navigus biopsy system (Medtronic) and the Leksell stereotactic frame (Elekta Ltd., Stockholm, Sweden). Methods: In the first experimental setup, we compared target accuracy of the Stealth Autoguide to the Navigus system by using phantom heads filled with gelatin to simulate the brain tissue. In the second experimental setup, we inserted SEEG electrodes to targets within cadaveric heads in a simulated operating room environment. Results: Using a homogeneous gelatin-filled phantom 3D reconstruction of a human head, we found that using the Stealth Autoguide system, while maintaining accuracy, was faster to use than the Navigus system. In our simulated operating room environment using nonliving human cadaveric heads, we found the accuracy of the Stealth Autoguide robotic device to be comparable to that of the Leksell frame. Discussion/Conclusion: These results compare the use of the Stealth Autoguide robotic guidance system with commonly used stereotactic devices, and this is the first study to compare its use and accuracy with the Leksell frame. These findings provide mounting evidence that Stealth Autoguide will have potential clinical uses in various stereotactic neurosurgical procedures.

Langue d'origineEnglish
Pages (de-à)343-350
Nombre de pages8
JournalStereotactic and Functional Neurosurgery
Volume99
Numéro de publication4
DOI
Statut de publicationPublished - juill. 1 2021

Note bibliographique

Funding Information:
The authors would like to acknowledge the support of the Department of Diagnostic Radiology Research Office (Dalhousie University, Halifax, NS, Canada) with image acquisition, SurgiOne Medical (Burlington, ON, Canada) and AdTech Medical (Oak Creek, WI, USA) for providing depth electrodes, and Medtronic for providing the use of the Navigus biopsy kits.

Publisher Copyright:
© 2021 S. Karger AG, Basel. Copyright: All rights reserved.