Abstract
We found total protein estimates in patients with monoclonal gammopathies to be erroneously low when using the Cobas-BioR centrifugal analyzer. This problem occurred only when a serum-water blank was used. This probably results from the precipitation of these proteins under conditions of low ionic strength resulting in high blank readings. The problem can be avoided if a serum-saline blank is used.
Original language | English |
---|---|
Pages (from-to) | 233-235 |
Number of pages | 3 |
Journal | Clinical Biochemistry |
Volume | 17 |
Issue number | 4 |
DOIs | |
Publication status | Published - Aug 1984 |
Externally published | Yes |