Artefactual decrease in total protein concentration in patients with monoclonal gammopathies: A method-dependent error

Karen J. Whitlow, Amin A. Nanji, Morris R. Pudek

Résultat de recherche: Articleexamen par les pairs

3 Citations (Scopus)

Résumé

We found total protein estimates in patients with monoclonal gammopathies to be erroneously low when using the Cobas-BioR centrifugal analyzer. This problem occurred only when a serum-water blank was used. This probably results from the precipitation of these proteins under conditions of low ionic strength resulting in high blank readings. The problem can be avoided if a serum-saline blank is used.