Fabrication and performance of 40-60 MHz annular arrays

Jeremy A. Brown, Christine E. Morton, Geoffrey R. Lockwood

Research output: Contribution to journalConference articlepeer-review

3 Citations (Scopus)

Abstract

The fabrication and performance of miniature high frequency (40-60 MHz) annular arrays is described. The arrays were fabricated using photolithography to define the electrode pattern, and ultrasonic wire bonding to make electrical contacts. The transducers were 2 mm in diameter with seven equal area elements and a centre frequency of 50 MHz. A -6 dB two-way beam width of 75 microns was found at f/2. The radiation pattern decreased smoothly to less than -60 dB at a distance of 550 microns.

Original languageEnglish
Pages (from-to)869-872
Number of pages4
JournalProceedings - IEEE Ultrasonics Symposium
Volume1
Publication statusPublished - 2003
Externally publishedYes
Event2003 IEEE Ultrasonics Symposium - Proceedings - Honolulu, HI, United States
Duration: Oct 5 2003Oct 8 2003

ASJC Scopus Subject Areas

  • Acoustics and Ultrasonics

Fingerprint

Dive into the research topics of 'Fabrication and performance of 40-60 MHz annular arrays'. Together they form a unique fingerprint.

Cite this